Rohde & Schwarz CMN50HF30 probe, Rogowski Current Probe, 1.5 Hz to 50 MHz, 6 kA

The Rohde & Schwarz CMN50HF30 probe is a Rogowski current probe for currents up to 6 kA, with a 1.5 Hz to 50 MHz band. Lost in HCM offers it for rental or purchase in Vietnam.

SKU: LIH-CMN50HF30
Category:

Description

Some currents are simply too large for a clamp. The Rohde & Schwarz CMN50HF30 probe is a Rogowski current probe. So it uses a thin, flexible coil that wraps around the conductor. An integrator then turns the coil signal into a current waveform for your oscilloscope. It handles up to 6 kA from 1.5 Hz to 50 MHz. Lost in HCM offers it for rental and sale in Vietnam.

Rohde & Schwarz CMN50HF30 probe family photo: open black hard case with a thin black Rogowski coil cable, a grey integrator box and printed instructions in the lid

Listed data

Order number 1801.8250.02
Probe type Rogowski current probe
Bandwidth 1.5 Hz to 50 MHz
Maximum current 6 kA
Sensitivity 1 mV/A
Cable / coil length 2.5 m / 100 mm
Coil voltage 5 kVpk

Where 6 kA of headroom pays off

Within the CMN50HF set, this model carries the highest current. So the Rohde & Schwarz CMN50HF30 probe suits the largest events on a power test floor, such as these.

  • Short-circuit and fault tests on big power modules.
  • Inrush when large motors or transformers switch on.
  • Pulsed power, welding and surge test rigs.
  • Switching current on bus bars in traction and storage inverters.

Living with a 1 mV/A output

A wide range always costs signal level. At 1 mV/A, a 100 A ripple shows up as only 100 mV on screen. Therefore set the channel scaling with care, and choose a smaller sibling when your currents stay low. Also, conductor position and nearby cables can shift a Rogowski reading. So follow the Rohde & Schwarz setup notes.

The coil also has a clear limit. Because its band begins at 1.5 Hz, it does not see steady DC. Instead it follows the changing part of a current, such as edges, ripple and pulses.

Buying or renting the Rohde & Schwarz CMN50HF30 probe

Many teams need a 6 kA coil for one test campaign, so a rental fits well. Others run fault tests every month and would rather own one. In both cases, request a quote from Lost in HCM with your scope model, peak and RMS current, conductor size and project dates. You can also browse other oscilloscope probes here, read the Rohde & Schwarz current probe page, or contact SMK Vietnam.

CMN50HF30 questions

Can the CMN50HF30 measure DC current?

No. Its listed band starts at 1.5 Hz, so it reads changing current only. For DC work, look at a clamp-style current probe instead.

Which oscilloscope input does it need?

The listing asks for a scope input that suits the probe output. Confirm the termination and scaling with Lost in HCM before you book.

How does it differ from the CMN50HF15?

Both share the 50 MHz top end, the 2.5 m cable and the 100 mm coil. However, the Rohde & Schwarz CMN50HF30 probe doubles the range to 6 kA. It also halves the output to 1 mV/A and reaches down to 1.5 Hz.

Additional information

Brand

Rohde & Schwarz

Supplied by

Lost in HCM (lostinhcm.com)

Reviews

There are no reviews yet.

Be the first to review “Rohde & Schwarz CMN50HF30 probe, Rogowski Current Probe, 1.5 Hz to 50 MHz, 6 kA”

Your email address will not be published. Required fields are marked *